HM-JACK arc

Automatic analyser to detect the fecal occult blood.
Small amount of stool and only few minutes will give the very proper and exact result allowing the doctors to summarise the possible status disease of the different patients

HM-JACK arc
Full automation:

Improves lab workflow and reduces operator minimum time for screening routine


User-friendly SW and LCD touchscreen:

Enables simple and fast analysis set-up, monitoring and report management


New rack system:

Increases throughput speed


Quantitative measurement:

Guarantees high reliability of the results


Measurement range 7-400 ng/ml:

Improves repeatability at lower end


Adjustable Cut-Off:

Allows flexibility for screening program


7 point Master Curve/2 point Calibration:

Guarantees precision and accuracy of the results


Quantitative controls:

Allows traceability and validation of the results


hmj-top-img-circle
Extel "HemoAuto HS"

High sensitivity latex:
improves repeatability at cut-off value.

Optimal concentration:
maximises the integrating sphere nephelometer.

This dedicated reagent features high sensitivity latex for integrating sphere turbidimetric assay applied to faecal occult blood analysis. Its new formula improves repeatability both at cut off value and lower end, enabling to achieve superior stability and quick quantification

dep ok

IST (scattered light/trasmitted light)                                      IST: Integrated Sphere Turbidimetry                

(1) Principle
(2) Trasmitted light
(3) Antigen
(4) Scattered light
(5) Light source
(6) Scattered light
(7) Trasmitted light

HemoAutoMC collection picker
  • 2 steps rotating scrubbing method: improves sampling accuracy.
  • Hexagonal hole: guarantees stability and solubility of the sample on board.
  • Sight glass: allows checking the correct collection of the sample.
  • Multifunctional integrated seal: allows better sample management and prevents leaking.
  • Over-sticker: prevents contamination of the piercing portion.
  • Internal and External Barcode: ensures traceability of samples and accuracy in QC procedure.

While the simple and ingenious design of this new collection picker enables an efficient and accurate sampling procedure, its newly developed isolation vessel guarantees perfect preservation from contamination. Its exclusive features are fine-tuned to work in unison with reagent and analyser to further enhance the outstanding operability and accuracy of the system.

dep ok

Legenda:

  1. Confirmable to break a sticker. Sticker to open.
  2. Hexagonal hole assures to retain the feces.
  3. Screw-in picking stick will reduce the buffer sperad on removal of feces picking stick.
  4. Sight glass of collecting sample.
  5. After collection of the sample.
  6. Before collection of the sample.
  7. Portion of piercing.
  8. Excessive feces to be retained a sight glass of collection sample.
  9. Prevent contamination with overlabeled (Keep sealed in unused state).
  10. Upgrading of molding technology brought implementation of thin-layer to pierce
Analysis method

Latex agglutination method


Principles of measurement

Integrated spherical nephelometry (turbidity) determination


Through put

200 samples/hour


Volume of test sample

10 ~ 20µL


Volume of reagent

Latex: 90µL Buffer : 190µL


Sample table

80 test sample ( Rack method )


Test sample tube

Exclusive feces pickers (HemoAuto MC)


Pipette for test sample

Direct insertion into feces pickers


Pipette for reagents

Pressed off nozzle system with buffer solution


Reaction table

Capacity for reaction cells placed: 40 cells


Reaction time

4.5 minutes


Supply of reaction cells

Automatic supply and discharge


Total number of reaction cells

80 cells (two cassettes with 40 reaction cells each)


Reaction temperature

Room temperature


Light source

Halogen lamp 5V 9W


Detector

Silicon photodiode


Calibration of concentration

Seven-point calibration curve


Calibration curve

Master-curve system, two-point correction


Monitor

Color Liquid-crystal monitor ( Touch panel )


Printer

32-figures thermal printer


Power supply

AC 100V ±10V or AC 220V±10V


Power consumption

300 VA


SizesCalibration of concentration

H 500 × W 600 × D 610 mm


Weight

56 kg


External output

RS232C